Clarity INT9 A/D Datasheet

Browse online or download Datasheet for Unknown Clarity INT9 A/D. Clarity INT9 A/D Datasheet User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Getting Started

ClarityGetting StartedClarity ENGCode/Rev.: M006/50BDate: 11. 9. 2013Phone: +420 251 013 400 DataApex Ltd.Fax: +420 251 013 401 Petrzilkova 2583/13cla

Page 2

Getting Started 2 Installation2.3 Hardware InstallationThe following chapters describe the installation of a protective Hardwarekey (dongle) and the b

Page 3 - Contents

2 Installation Clarity2.3.2 The INT9 CardIn Windows 8/7/Vista/XP the drivers will be installed automatically duringthe installation of Clarity.l Insta

Page 4 - Table of Contents Clarity

Getting Started 2 Installation2.4 Device setup and wiringThe wiring strongly depends on the particular station. Clarity stationpackages which use an A

Page 5

2 Installation Clarity2.4.2 ChromatographConnect cables according to one of the diagrams in Fig 4 on pg 7. Use asymmetrical connection only if you are

Page 6

Getting Started 2 InstallationFor easier modification of the wiring we supply the SV9 Terminal board(p/n SV9) with screw contacts for INT9, U-PAD2 or

Page 7 - 1 Brief Description

2 Installation Clarity2.5 Clarity ConfigurationNote: Clarity is automatically preconfigured for analogue acquisition with nocontrol modules. However,

Page 8 - 2 Installation

Getting Started 2 InstallationNote: If necessary, add further equipment to the list on the left using the Addbutton in the lower left-hand section of

Page 9

Getting Started 3 Qualification procedures3 Qualification proceduresThe quality of analytical data is an important issue in many laboratories.One of t

Page 10 - 2.3 Hardware Installation

3 Qualification procedures ClarityFig 8: IQ windowl If the Installation Qualification fails, it is recommended to uninstall andthen re-install Clarity

Page 11 - 2.3.2 The INT9 Card

Getting Started 3 Qualification proceduresThe Validation kit (p/n: CVK), as well as SST Extension (p/n: A22), canbe purchased separately. The validati

Page 12 - 2.4 Device setup and wiring

Clarity®, DataApex®and®are trademarks of DataApex Ltd. Microsoft®and WindowsTMaretrademarks of Microsoft Corporation.DataApex reserves the right to ma

Page 13 - 2.4.2 Chromatograph

Getting Started 4 Program structure and control4 Program structure and controlClarity has a hierarchic structure. After start-up, the main Clarity win

Page 14 - 2.4.3 Autosampler

4 Program structure and control ClarityAll dialogs relevant for the performance of actions in the Instrumentwindow can be easily accessed from the Ins

Page 15 - 2.5 Clarity Configuration

Getting Started 5 Tour through the Clarity station5 Tour through the Clarity stationThe following two sections will show you step by step,how to perfo

Page 16

5 Tour through the Clarity station Clarity5.1 Running the Single AnalysisThere is a simple project aimed on basic functions prepared for Instrument2.

Page 17 - 3 Qualification procedures

Getting Started 5 Tour through the Clarity station5.1.2 Single Analysis dialogUse the Single Analysis button in the Instrument window to open theSingl

Page 18

5 Tour through the Clarity station Clarityl Run the analysis by clicking the Run button . The Single Analysis dialogwill close now, but if you open it

Page 19

Getting Started 5 Tour through the Clarity station5.1.3 Data Acquisition windowl In the Instrument window, look at the Status line (see Fig 10 on pg 1

Page 20

5 Tour through the Clarity station Clarityl The Chromatogram window opens because the station is set to do so.These settings are available in the Inst

Page 21

Getting Started 5 Tour through the Clarity station5.1.4 Chromatogram windowl The Chromatogram window can also be opened manually by clicking onthe Chr

Page 22

5 Tour through the Clarity station Claritycorner of the graph (the active signal is set in bold), from the icons in theOverlay toolbar (the active sig

Page 23 - 5.1.1 Instrument window

Getting Started Table of ContentsContents1 Brief Description 11.1 Hardware and software requirements 12 Installation 22.1 Multilingual Clarity 22.2 So

Page 24 - 5.1.2 Single Analysis dialog

Getting Started 5 Tour through the Clarity station5.2 Running the Sequence measurementSequence operation allows automated measurement of large numbers

Page 25

5 Tour through the Clarity station Claritycalibration (or its recalibration, if there already were any data in thecalibration).l The Method Name colum

Page 26 - 5.1.3 Data Acquisition window

Getting Started 5 Tour through the Clarity station(no chromatogram saved) or Stop button (resulting chromatogram willbe saved).- 26 -

Page 27

5 Tour through the Clarity station Clarity5.2.2 Calibration windowlUse the Calibration button in the Instrument window to open theCalibration window.N

Page 28 - 5.1.4 Chromatogram window

Getting Started 5 Tour through the Clarity stationFig 17: Calibration window - loaded standardl Set the calibration level number to 1 on the Current L

Page 29

5 Tour through the Clarity station Clarityl Set the third calibration level accordingly using the STD 3.PRM file andAmount of 3.0 and the fourth level

Page 30 - 5.2.1 Sequence window

Getting Started 5 Tour through the Clarity station5.2.4 Linking the calibration to the methodIn case of a large number of chromatograms, linking the c

Page 31

5 Tour through the Clarity station ClarityMark all files with the names SAMPLE_VIAL_X-Y in the DATA directory tobe reprocessed, check the Reprocess by

Page 32

Getting Started 6 Connecting Autosamplers6 Connecting AutosamplersThis chapter describes wiring of the autosamplers. The configurationvaries depending

Page 33 - 5.2.2 Calibration window

6 Connecting Autosamplers ClarityVariant A - started by closing the contactsThe first diagram shows the wiring of an autosampler that will initiate th

Page 34

Table of Contents Clarity7.2.5 DEMO (in the window header) 437.2.6 User accounts file load error 437.3 Problems upon collection of data 447.3.1 Data A

Page 35

Getting Started 6 Connecting AutosamplersVariant B - started by opening the contactsIn the second scheme there is autosampler wiring that conversely w

Page 36 - Ctrl or Shift key

6 Connecting Autosamplers Clarity6.2 AS + LC set - Active SequenceIn LC systems, the autosampler typically governs the timings. Theeventual pump gradi

Page 37

Getting Started 6 Connecting Autosamplers6.3 AS - Passive Sequence (GC or LC)The autosampler used in the Clarity Passive Sequence (both GC and LCsets)

Page 38 - 6 Connecting Autosamplers

6 Connecting Autosamplers Clarity6.4 Auto Sampler - Active Sequence + AS Control + A/D cardWhen using the optional AS Control (p/n A26) module, all co

Page 39

Getting Started 6 Connecting Autosamplers6.5 AS - Active Sequence + AS Control + digital acquisitionWhen using an optional AS Control module in combin

Page 40

Getting Started 7 Troubleshooting7 TroubleshootingIf you will not find your answers here, use the www.dataapex.comwww.dataapex.com website where the S

Page 41

7 Troubleshooting Clarity7.1 Locate your problem:When troubles occur, the fastest way to find a solution for it is to search forit in the following in

Page 42

Getting Started 7 Troubleshooting7.2 Problems at station start7.2.1 DEMO - missing HW KeylockThe Rockey (or Sentinel) protective key (dongle) must be

Page 43

7 Troubleshooting Clarity7.2.2 DEMO - Keylock test failed▌ There is a DEMO - Keylock test failed error message in the Clarity window.Fig 25: DEMO - Ke

Page 44

Getting Started 7 Troubleshooting7.2.5 DEMO (in the window header)▌ The DEMO inscription is displayed in the header of the Clarity window.Fig 26: Clar

Page 45 - 7 Troubleshooting

Getting Started Table of ContentsTo facilitate the orientation in the Getting Started manual and Clarity chromatography station,different fonts are us

Page 46 - 7.1 Locate your problem:

7 Troubleshooting Clarity7.3 Problems upon collection of data7.3.1 Data Acquisition - non-functionalGray icon with the heading "Disabled" an

Page 47 - 7.2 Problems at station start

Getting Started 7 Troubleshooting▌ c) The detector is not allocated to the Clarity Instrument:Fig 30: Detector not allocated to the InstrumentDescript

Page 48 - 7.2.4 DEMO - Trial Expired

7 Troubleshooting Clarity7.3.2 Data Acquisition - Simulated▌ The title "Simulated" is displayed on the Data Acquisition window.Fig 32: Data

Page 49

Getting Started 7 Troubleshooting7.3.3 Other Error MessagesYou can find the description of other Error Messages and possibleproblems and solutions for

Page 50

7 Troubleshooting Clarity7.4 Hardware Key7.4.1 Rockey USB dongle (re)installationFig 33: Rockey USB donglel Install Clarity software.Caution: Be sure

Page 51

Getting Started 7 Troubleshooting7.4.2 Rockey LPT dongle (re)installationFig 34: Rockey LPT dongleIf you encounter problems with printer port Rockey h

Page 52

7 Troubleshooting Clarity7.4.3 Sentinel dongle installationUntil Clarity version 2.4.1 the dongle had to be installed and present inthe PC only when u

Page 53 - 7.3.3 Other Error Messages

Getting Started 7 Troubleshooting7.5 System Files (systeminfo.txt file)The C:\CLARITY\SYSTEMINFO.TXT file contains valuable diagnosticinformation. Its

Page 54 - 7.4 Hardware Key

7 Troubleshooting Claritystation, number of Instruments allowed, extensions available, purchasedcontrol modules, type and serial number of the hardwar

Page 56 - Sentinel Installer

Getting Started 1 Brief Description1 Brief DescriptionClarity chromatography station is an effective tool for the acquisition,processing and evaluatio

Page 57

Getting Started 2 Installation2 InstallationVerify that the package is complete (packing list).Caution: Install Clarity before inserting any devices (

Page 58 - 7.6 Sleep Mode

2 Installation ClarityNote: If you are only evaluating the program, do not enter the User Code andpress the Skip button.l During installation of the d