Clarity - INT9 A/D converter Installation
11
2.4.1 Measuring on multiple Instruments
Rather than dragging the entire INT9 A/D
Card Driver icon from the Setup Control
Modules list (see Fig. 5), drag the
individual detector signals (TCD, FID in this
case) to separate Instruments . The
Instrument to which the particular signal is
connected is listed to the right of the
detector name in the Setup Control Modules
section.
Fig. 5. System Configuration – multiple Instruments
Note: Each independent Clarity Instrument must have
separate Start and Ready pins configured according
to the actual wiring.